T1 - A Novel Bioresorbable Polymer Paclitaxel-Eluting Stent for the Treatment of Single and Multivessel Coronary Disease. Primary Results of the COSTAR (Cobalt Chromium Stent With Antiproliferative for Restenosis) II Study
AU - Krucoff, Mitchell W.
AU - Kereiakes, Dean J.
AU - Petersen, John L.
AU - Mehran, Roxana
AU - Hasselblad, Vic
AU - Lansky, Alexandra J.
AU - Fitzgerald, Peter J.
AU - Garg, Jyotsna
AU - Turco, Mark A.
AU - Simonton, Charles A.
AU - Verheye, Stefan
AU - Dubois, Christophe L.
AU - Gammon, Roger
AU - Batchelor, Wayne B.
AU - O'Shaughnessy, Charles D.
AU - Hermiller, James B.
AU - Schofer, Joachim
AU - Buchbinder, Maurice
AU - Wijns, William
N1 - Funding Information:
Supported by a research grant from Conor MedSystems. For full author disclosures, please see the end of this paper.
PY - 2008/4/22
Y1 - 2008/4/22
N2 - Objectives: The aim was to compare safety and effectiveness of the CoStar drug-eluting stent (DES) (Conor MedSystems, Menlo Park, California) with those of the Taxus DES (Boston Scientific, Maple Grove, Minnesota) in de novo single- and multivessel percutaneous coronary intervention (PCI). Background: Paclitaxel elution from a stent coated with biostable polymer (Taxus) reduces restenosis after PCI. The CoStar DES is a novel stent with laser-cut reservoirs containing bioresorable polymer loaded to elute 10 μg paclitaxel/30 days. Methods: Patients undergoing PCI for a single target lesion per vessel in up to 3 native epicardial vessels were randomly assigned 3:2 to CoStar or Taxus. Primary end point was 8-month major adverse cardiac events (MACE), defined as adjudicated death, myocardial infarction (MI), or clinically driven target vessel revascularization (TVR). Protocol-specified 9-month angiographic follow-up included 457 vessels in 286 patients. Results: Of the 1,700 patients enrolled, 1,675 (98.5%) were evaluable (CoStar = 989; Taxus = 686), including 1,330 (79%) single-vessel and 345 (21%) multivessel PCI. The MACE rate at 8 months was 11.0% for CoStar versus 6.9% for Taxus (p < 0.005), including adjudicated death (0.5% vs. 0.7%, respectively), MI (3.4% vs. 2.4%, respectively), and TVR (8.1% vs. 4.3%, respectively). Per-vessel 9-month in-segment late loss was 0.49 mm with CoStar and 0.18 mm with Taxus (p < 0.0001). Findings were consistent across pre-specified subgroups. Conclusions: The CoStar DES is not noninferior to the Taxus DES based on per-patient clinical and per-vessel angiographic analyses. The relative benefit of Taxus is primarily attributable to reduction in TVR. Follow-up to 9 months showed no apparent difference in death, MI, or stent thrombosis rates.
AB - Objectives: The aim was to compare safety and effectiveness of the CoStar drug-eluting stent (DES) (Conor MedSystems, Menlo Park, California) with those of the Taxus DES (Boston Scientific, Maple Grove, Minnesota) in de novo single- and multivessel percutaneous coronary intervention (PCI). Background: Paclitaxel elution from a stent coated with biostable polymer (Taxus) reduces restenosis after PCI. The CoStar DES is a novel stent with laser-cut reservoirs containing bioresorable polymer loaded to elute 10 μg paclitaxel/30 days. Methods: Patients undergoing PCI for a single target lesion per vessel in up to 3 native epicardial vessels were randomly assigned 3:2 to CoStar or Taxus. Primary end point was 8-month major adverse cardiac events (MACE), defined as adjudicated death, myocardial infarction (MI), or clinically driven target vessel revascularization (TVR). Protocol-specified 9-month angiographic follow-up included 457 vessels in 286 patients. Results: Of the 1,700 patients enrolled, 1,675 (98.5%) were evaluable (CoStar = 989; Taxus = 686), including 1,330 (79%) single-vessel and 345 (21%) multivessel PCI. The MACE rate at 8 months was 11.0% for CoStar versus 6.9% for Taxus (p < 0.005), including adjudicated death (0.5% vs. 0.7%, respectively), MI (3.4% vs. 2.4%, respectively), and TVR (8.1% vs. 4.3%, respectively). Per-vessel 9-month in-segment late loss was 0.49 mm with CoStar and 0.18 mm with Taxus (p < 0.0001). Findings were consistent across pre-specified subgroups. Conclusions: The CoStar DES is not noninferior to the Taxus DES based on per-patient clinical and per-vessel angiographic analyses. The relative benefit of Taxus is primarily attributable to reduction in TVR. Follow-up to 9 months showed no apparent difference in death, MI, or stent thrombosis rates.
UR - https://www.scopus.com/pages/publications/41949127381
U2 - 10.1016/j.jacc.2008.01.020
DO - 10.1016/j.jacc.2008.01.020
M3 - Article
C2 - 18420096
AN - SCOPUS:41949127381
SN - 0735-1097
VL - 51
SP - 1543
EP - 1552
JO - Journal of the American College of Cardiology
JF - Journal of the American College of Cardiology
IS - 16
ER -