Endocarditis and incomplete endothelialization: 12 years after Amplatzer septal occluder deployment

  • Allan K. Nguyen
  • , Brian A. Palafox
  • , Joanne P. Starr
  • , Richard N. Gates
  • , Farbouch Berdjis

Research output: Contribution to journalArticlepeer-review

21 Scopus citations

Abstract

A 4-year-old boy had a 15-mm atrial septal defect repaired percutaneously with use of an Amplatzer Septal Occluder. At age 16 years, he presented with a week’s history of fever, chills, dyspnea, fatigue, and malaise. Cultures grew methicillin-sensitive Staphylococcus aureus. A transesophageal echocardiogram showed a 1.25 × 1.5-cm pedunculated mass on the left aspect of the atrial septum just superior to the mitral valve, and a smaller vegetation on the right inferior medial aspect of the septum. At surgery, visual examination of both sides of the septum revealed granulation tissue, the pedunculated mass, the small vegetation, and exposed metal wires that suggested incomplete endothelialization of the occluder. We removed the occluder and patched the septal defect. The patient returned to full activity after 4 months and was asymptomatic 3 years postoperatively. Our report reinforces the need for further investigation into prosthetic device endothelialization, endocarditis prophylaxis, and recommended levels of physical activity in patients whose devices might be incompletely endothelialized. In addition to reporting our patient’s case, we review the medical literature on this topic.

Original languageEnglish
Pages (from-to)227-231
Number of pages5
JournalTexas Heart Institute Journal
Volume43
Issue number3
DOIs
StatePublished - Jun 2016

Keywords

  • Bacterial/etiology/pathology/ prevention & control
  • Biocompatible materials/metabolism
  • Cardiac catheterization/instrumentation
  • Congenital/ therapy
  • Endocarditis
  • Endothelial cells/metabolism
  • Heart defects
  • Heart valve prosthesis implantation/adverse effects/instrumentation/ methods
  • Prosthesis-related infections/etiology/therapy
  • Septal occluder device/adverse effects
  • Time factors
  • Treatment failure

Fingerprint

Dive into the research topics of 'Endocarditis and incomplete endothelialization: 12 years after Amplatzer septal occluder deployment'. Together they form a unique fingerprint.

Cite this