Skip to main navigation Skip to search Skip to main content

Insertion and recovery of a new retrievable vena caval filter work in progress

  • George P. Teitelbaum
  • , Mark McGurrin
  • , Barbara L. Davies
  • , Michael J. Moritz
  • , R. Anthony Carabasi
  • , Bruce E. Jarrell
  • , George W. English

Research output: Contribution to journalArticlepeer-review

11 Scopus citations

Abstract

A new retrievable inferior vena cava (IVC) filter was tested in nine pigs. Insertion was through a 14 French sheath using both the femoral and jugular approaches. All insertions were successful, and there was a 100% postinsertion IVC patency rate (8/8 pigs at one week and 1/1 pig at one month). Addition of an apical hook to the filter design allowed transjugular retrieval of two filters at one week postinsertion. Three of nine filters migrated to the upper IVC. The filter’s design allows paraxial blood flow despite trapped thrombus and inhibits filter tilting. In vitro, the filter captured 95% to 100% of 5 × 5 mm clots. If problems with migration can be solved, the new filter may provide effective short- and long-term prophylaxis against pulmonary embolism.

Original languageEnglish
Pages (from-to)527-533
Number of pages7
JournalInvestigative Radiology
Volume23
Issue number7
DOIs
StatePublished - Jul 1988

Keywords

  • Embolism
  • Filters
  • Pulmonary
  • Venae cavae

Fingerprint

Dive into the research topics of 'Insertion and recovery of a new retrievable vena caval filter work in progress'. Together they form a unique fingerprint.

Cite this