Skip to main navigation Skip to search Skip to main content

MR imaging in patients with intracranial aneurysm clips

  • R. L. Becker
  • , J. F. Norfray
  • , G. P. Teitelbaum
  • , W. G. Bradley
  • , J. B. Jacobs
  • , L. Wacaser
  • , R. L. Rieman

Research output: Contribution to journalArticlepeer-review

49 Scopus citations

Abstract

Four patients with intracranial aneurysm clips made from a variety of alloys were studied without incidence by MR imaging at field strengths ranging from 0.35 to 0.6 T. Knowledge of the type of alloy used in the manufacturing of an aneurysm clip is important in determining whether the clip will or will not deflect in a magnetic field. Ferromagnetic clips show deflection and torque in a magnetic field and have the potential to dislodge from the aneurysm. Nonferromagnetic or weakly ferromagnetic aneurysm clips such as the Sugita (Elgiloy), Yasargil (316 LVM stainless steel), Heifetz (Elgiloy), Yasargil (Phynox), and Vari-Angle McFadden (MP35N) do not deflect or deflect weakly in the magnetic field and therefore would not be expected to dislodge during MR. The option of imaging many patients with intracranial aneurysm clips with MR extends the usefulness of the technique to a previously excluded population.

Original languageEnglish
Pages (from-to)885-890
Number of pages6
JournalAmerican Journal of Neuroradiology
Volume9
Issue number5
StatePublished - 1988

Fingerprint

Dive into the research topics of 'MR imaging in patients with intracranial aneurysm clips'. Together they form a unique fingerprint.