Skip to main navigation Skip to search Skip to main content

Strut interlocking of Titanium Greenfield vena cava filters and its effect on clot capturing: An in vitro study

Research output: Contribution to journal › Article › peer-review

15 Scopus citations

Abstract

Purpose: To determine the influence of filter leg-interlocking on filtering efficiency of titanium Greenfield inferior vena cava filters in an in vitro model. Methods: Titanium Greenfield filters (TGF) were placed in an inferior vena cava (IVC) flow phantom with no interlocking legs, or with one or two pairs of legs interlocked. Clot emboli of varying sizes were introduced into the stream of flow, and the frequency of successful captures by the deformed filters was recorded. Results: Interlocking filter legs diminished the filtering efficiency of the TGF by up to 80% with the flow phantom in a horizontal position; no such compromise existed when the phantom was in a vertical position. Conclusion: When leg interlocking is verified radiographically, the filtering efficiency of the TGF is reduced, and placement of a second IVC filter may be indicated.

Original languageEnglish
Pages (from-to)204-206
Number of pages3
JournalCardioVascular and Interventional Radiology
Volume17
Issue number4
DOIs
StatePublished - Jul 1994

Keywords

  • Filter, complications
  • Pulmonary embolism

Fingerprint

Dive into the research topics of 'Strut interlocking of Titanium Greenfield vena cava filters and its effect on clot capturing: An in vitro study'. Together they form a unique fingerprint.

Cite this